100L Refrigerated Heating Baths Circulators 100L Refrigerated Heating Baths Circulators 100L Refrigerated Heating Baths Circulators
100L Refrigerated Heating Baths Circulators 100L Refrigerated Heating Baths Circulators 100L Refrigerated Heating Baths Circulators

100L Refrigerated Heating Baths Circulators

LGD-100/20 LGD-100/40 LGD-100/80
LABOAO LGD series refrigerated circulating baths provides outstanding performance for routine and sophisticated applications requiring accurate temperature control in the range of -80°C/-40°C/-20°C to 200°C.
Request QuotationChat Now
  • Whatsapp

    WhatsApp Chat

  • WeChat

    WeChat Chat

Download

DESCRIPTION

Recirculating Chiller Heater including: cooling function, heating function circulation system. It is composed of refrigeration unit, heating unit, circulation system and control system.

Refrigeration part is mechanically refrigerated by a fully enclosed compressor, composed of plate heat exchangers, expansion valves, and pressure protectors to meet the cooling requirements.

Heating part is the use of electric heating tubes to heat the circulating fluid. And through the built-in circulating pump and the external circulating pipeline, the refrigerant or heat medium is transported to the interlayer of the supporting equipment, which indirectly cools and heats the materials in the reactor.

During external circulation, connect the upper outlet pipe of the product to the lower inlet pipe of the kettle body, and the circulating fluid will return to the lower return port of the product through the pipe from the upper circulation port to form a complete circulation space.

FEATURES

1. The temperature accuracy is generally ±1°C, and an external sensor can be connected (the temperature difference accuracy is generally ±0.5°C).

2. The maximum temperature limit to prevent the equipment from losing control [If the maximum temperature of the equipment is exceeded, the equipment will automatically power off.

3. Has power supply protection.

4. Compressor overheating protection, the compressor will stop working when the compressor withstand limit is reached.

5. Water cooling has water flow and low water pressure protection.

6. All compressors are imported compressors (France, Denmark).

SPECIFICATIONS

Model

LGD-100/20

LGD-100/40

LGD-100/80

Matched reactor

80L-100L

80L-100L

80L-100L

Tank Volume(L)

30

30

30

Heating volume(L)

3

3

3

Heating tank size(mm)

90×480

90×480

90×480

Oil tank size(mm)

450*150*450

450*150*450

450*150*450

Working temperature range()

-20 to +200

-40 to +200

-80 to 200

Ambient temperature()

≤25

≤25

≤25

Lowest temperature (No load)

-22

-42

-82

Power

380V±10% 50HZ

380V±10% 50HZ

380V±10% 50HZ

Heating power(W)

9000

9000

9000

Cooling power(W)

11240-2100

10280-2565

105700-3270

Temperature accuracy()

±0.1

±0.1

±0.1

Sensor

PT100

PT100

PT100

Medium quality degree

500C.S.T

500C.S.T

500C.S.T

Circulating pump power(w)

250

750

750

Lift head(m)

15

25

25

Flow(L/min)

30

35

35